Operation of silicon photomultipliers in a dilution refrigerator down to 9.4 mK towards a cryogenic cosmic ray muon veto system
D. E. Zmeev · 2025
Read Operation of silicon photomultipliers in a dilution refrigerator down to 9.4 mK towards a cryogenic cosmic ray muon veto system by D. E. Zmeev (2025) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/arxiv-2512-16769.txt.
Introduction
The QUEST-DMC team tested a silicon photomultiplier sensor at extreme cold, down to 9.4 thousandths of a degree above zero, inside a dilution refrigerator. Such sensors could form a veto that flags cosmic-ray muons inside cold dark matter searches, cutting false signals.
The sensor, an FBK NUV-HD-cryo device, was checked for single-photon response, gain, dark count rate, and linked noise as voltage changed. The team also coupled it to a scintillator inside the fridge and saw events in line with cosmic-ray muons passing through. Both results are first steps toward a working cryogenic muon veto for the QUEST-DMC dark matter search.
Catalog record
| CREATOR | D. E. Zmeev |
|---|---|
| YEAR | 2025 |
| COLLECTION | PAPERS |
| FORMAT | TXT |
| FILE SIZE | 49 KB |
| VERSION | 1.0 |
| LICENSE | http://creativecommons.org/licenses/by/4.0/ |
| RIGHTS | CC BY |
| SOURCE | arXiv |
Source: https://arxiv.org/abs/2512-16769
Browse the full library · About the Tower of Babel
Full text
The opening of the work, from the library's hosted text:
Operation of silicon photomultipliers in a dilution refrigerator down to 9.4 mK towards a cryogenic cosmic ray muon veto system D. E. Zmeev Abstract We report the characterisation of a FBK NUV-HD-cryo silicon photomultiplier (SiPM) sensor operated in an 9.4 ± \pm 0.2 mK environment inside a dilution refrigerator, towards the development of a cryogenic cosmic ray muon veto system to be operated internal to a dilution refrigerator required for low background experiments such as the QUEST-DMC dark matter search experiment. We characterise the single photon response and the gain (the charge produced per detected photon), the dark count noise rate, and correlated noise contributions as a function of operating voltage. This paper also reports first proof of concept measurements of using a SiPM directly coupled to scintillator in a 9.4 mK temperature environment, towards detecting candidate cosmic ray muon signals. 1 Introduction In recent years, silicon photomultipliers (SiPMs) have emerged as promising single photon detectors for use in particle physics, particularly in low background, rare-event search experiments such as direct dark matter detection experiments. Photomultiplier Tubes (PMTs) have long been the favoured technology for dark matter experiments, such as DEAP-3600 and LZ [ 6 , 5 ] , however compared to PMTs, SiPMs are cheaper to manufacture, require a lower operational voltage, and are insensitive to magnetic fields. In addition to their high gain and high photon detection efficiency, they can have a lower intrinsic radioactivity compared to PMTs. One drawback of SiPMs is that they generally exhibit higher intrinsic thermal noise than PMTs. However, since thermal noise decreases proportionally with temperature, it can be suppressed by operating the SiPMs at cryogenic temperatures. Whilst limited, there have been measurements showing that SiPMs can operate close to liquid helium (LHe) temperatures and below [ 9 , 16 , 26 ] , illustrating that there is a growing interest to demonstrate the viability of operating these sensors in sub-Kelvin conditions. As such, SiPMs are being explored by the QUEST-DMC collaboration as a potential photon sensor for use in a cryogenic cosmic ray muon veto system. The QUEST-DMC collaboration is developing, in one of its work packages, a direct dark matter search experiment projected to reach world-leading sensitivity to spin-dependent sub-GeV/c 2 mass dark matter using a superfluid helium-3 ( 3 He) target enclosed in a \sim 1 cm 3 bolometer box [ 7 , 11 , 12 ] . The QUEST-DMC collaboration is operating bolometers at Royal Holloway, University of London and Lancaster University, in ultra-low temperature dilution refrigerators fitted with an adiabatic demagnetisation stage capable of cooling 3 He to operating temperatures below 300 µK. Since QUEST-DMC is currently operated at sea level, and not underground like the majority of dark matter experiments, cosmic ray induced background interactions will be a limiting factor for QUEST-DMC’s dark matter search [ 8 ] . The cosmic ray muon veto system is currently in development; the baseline design makes use of a scintillator volume surrounding the bolometer cell at \sim 300 µK, where the exterior of this volume is coupled to the inner vacuum can which itself is coupled to the 10 mK stage of the dilution refrigerator, and is where the SiPM will be mounted. Therefore, 10 mK is the relevant temperature to characterise the SiPM in order to demonstrate the feasibility of this scheme. We are considering SiPMs over technologies traditionally deployed for photon detection in this temperature range due to the fact that they are large-area, commercially manufacturable, and simple to operate, compared with Superconducting Nanowire Single-Photon detectors (SNSPDs) which are much smaller, and Transition Edge Sensors (TESs) which have a relatively complex fabrication process. SiPM technology has previously been used in cryogenic cosmic ray muon veto systems, such as the NUCLEUS experiment [ 13 ] , which makes use of a plastic scintillator deployed in a dilution refrigerator at 850 mK. In this configuration, the plastic scintillator is instrumented with wavelength shifting fibres, and is coupled to a SiPM that is thermally coupled to the 300 K part of the cryostat. However, we are exploring the option of directly coupling SiPMs to the plastic scintillator itself, which would require the SiPM technology to be able to operate in \sim mK environments. Without amplifier electronics and in the absence of light, a single NUV-HD-cryo SiPM with a surface area of 12 \times 8 mm 2 dissipates \mathcal{O} (pW) of power whilst it is biased, based on the device’s properties measured at 77 K [ 14 ] . At this level of power dissipation, the heat load would be low enough such that a SiPM could be thermally coupled to a \mathcal{O} (10’s) mK temperature stage of a dilution refrigerator without disrupting the overall cooling power of the cryostat. In this paper, we show the first operation of a NUV-HD-cryo SiPM [ 14 ] in a dilution refrigerator, demonstrating the feasibility of deploying this technology in a cryogenic cosmic ray muon system for the QUEST-DMC experiment. The SiPM is attached to the mixing chamber plate of the refrigerator, reaching a base temperature of 9.4 \pm 0.2 mK. Section 2 outlines the experimental setup, the data acquisition, and data processing chain. Section 3 describes the characterisation analysis that was performed. Section 4 reports first proof-of-principle measurements of using a SiPM directly coupled to a scintillator, and operated inside a dilution refrigerator. 2 Experimental setup 2.1 Hardware We operate a single NUV-HD-cryo SiPM [ 14 ] of 12 \times 8 mm 2 inside a cryogen-free dilution refrigerator located at Royal Holloway, University of London. The dilution refrigerator is a modified Oxford Instruments Triton 200 [ 17 ] . A simplified schematic of the experimental setup is shown in Figure 1 . The SiPM is mounted onto a custom PCB that distributes bias voltage and housed inside a 7 cm (L) x 3.8 cm (W) x 1.4 cm (D) copper box, which is attached to a copper stand that is thermally coupled to the plate of the mixing chamber. The base temperature is measured using current sensing noise thermometry (CSNT) [ 21 ] . A CSNT sensor with resistance 2 m \Omega is mounted onto the ultra-low temperature plate, which is suspended from the mixing chamber plate using alumina tubes. The mixing chamber plate is thermally coupled to the ultra-low temperature plate via a flexible copper link. Whilst the data was acquired for this work, the base temperature as measured from the CSNT sensor on the ultra-low temperature plate, and thus, the mixing chamber plate, was T_{\mathrm{MC}} = 9.4 \pm 0.2 mK. Further detail on the fridge design can be found in [ 19 ] . To ensure that the characterisation results we obtain actually reflect the SiPM operating in a 9.4 mK environment, we waited at least three days after cool-down to the mixing chamber base temperature before acquiring any data for this work. This allows for the plastic (FR-4) PCB on which the SiPM is mounted to thermalise to the same temperature as the copper. As Figure 1 shows, all associated electronics are located outside of the dilution refrigerator at room temperature, including a custom-designed signal amplification board, providing a 130 V/V voltage gain. A Keysight E3649A provides the low-voltage supply required