MKS 390 Micro-Ion Gauge Performance After Exposure to High Pressures

B. Massett 1 , W. T. Shmayda 2 · 2025

Read MKS 390 Micro-Ion Gauge Performance After Exposure to High Pressures by B. Massett 1 , W. T. Shmayda 2 (2025) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/arxiv-2510-25980.txt.

Description

Brandon Massett and Walter Shmayda check whether a vacuum gauge still reads correctly after an accidental blast of high pressure. The device is the MKS 390 Micro-Ion Gauge, used to measure very low pressures. In some uses a gauge may face high pressure on purpose or by accident, which raises concern about later accuracy and about leaks.

Massett and Shmayda compare two gauges side by side, one exposed to ten thousand Torr and one kept below one thousand Torr. Torr is a unit of pressure tied to a millimeter of mercury. Differences between the readings stayed within the maker's stated uncertainty, which shows negligible harm from the exposure, and the exposed gauge kept its leak integrity.

Catalog record

CREATORB. Massett 1 , W. T. Shmayda 2
YEAR2025
COLLECTIONPAPERS
FORMATTXT
FILE SIZE12 KB
VERSION1.0
LICENSEhttp://creativecommons.org/licenses/by/4.0/
RIGHTSCC BY
SOURCEarXiv

Source: https://arxiv.org/abs/2510-25980

Browse the full library · About the Tower of Babel