Technical Note: Low-Dose Simulation for Grating-Based X-Ray Dark-Field Radiography Using a Virtually Decreased Irradiation Area

2025

Read Technical Note: Low-Dose Simulation for Grating-Based X-Ray Dark-Field Radiography Using a Virtually Decreased Irradiation Area (2025) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/arxiv-2508-15328.txt.

Introduction

Henriette Bast, Rafael Schick, Thomas Koehler, and Franz Pfeiffer validate a way to fake low-dose lung X-rays from full-dose ones, in a technical note from 2025. Dark-field radiography shows the fine structure of the lung's air sacs by tracking scattered X-rays. To study dose cuts safely, researchers simulate low-dose images by shrinking the irradiated area in software.

The team scanned a chest phantom at several real dose levels and compared true low-dose images with simulated ones. At a quarter of the standard dose, real images stayed clean. Simulated images differed by up to ten percent in signal and showed stripe artifacts at deep dose cuts. The authors show those stripes come from the algorithm itself, not from real low-dose scans.

Catalog record

YEAR2025
COLLECTIONPAPERS
FORMATTXT
FILE SIZE17 KB
VERSION1.0
LICENSEhttp://creativecommons.org/licenses/by/4.0/
RIGHTSCC BY
SOURCEarXiv

Source: https://arxiv.org/abs/2508-15328

Browse the full library · About the Tower of Babel

Full text

The opening of the work, from the library's hosted text:

Technical Note: Low-Dose Simulation for Grating-Based X-Ray Dark-Field Radiography Using a Virtually Decreased Irradiation Area Abstract Background: X-ray dark-field radiography uses small-angle scattering to visualize the structural integrity of lung alveoli. To study the influence of dose reduction on clinical dark-field radiographs, one can simulate low-dose images by virtually reducing the irradiated area. However, these simulations can exhibit stripe artifacts. Purpose: Validation of the low-dose simulation algorithm reported by Schick & Bast et al., PLoS ONE, 2024 in [ 1 ] . Furthermore, we want to demonstrate that stripe artifacts observed in simulated images at very low-dose levels are introduced by limitations of the algorithm and would not appear in actual low-dose dark-field images. Methods: Dark-field radiographs of an anthropomorphic chest phantom were acquired at different tube currents equaling different radiation doses. Based on the measurement with a high radiation dose, dark-field radiographs corresponding to lower radiation doses were simulated by virtually reducing the irradiated area. The simulated low-dose radiographs were evaluated by a quantitative comparison of the dark-field signal using different regions of interests. Results: Dark-field radiographs acquired at one quarter of the standard dose were artifact-free. The dark-field signal differed from the simulated radiographs by up to 10%. Algorithm-induced stripe artifacts decrease the image quality of the simulated radiographs. Conclusions: Virtually reducing the irradiation area is a simple approach to generate low-dose radiographs based on images acquired with scanning-based dark-field radiography. However, as the algorithm creates stripe artifacts in the dark-field images, particularly at higher dose reductions, that are not present in measured low-dose images, simulated images have reduced image quality compared to their measured counterparts. Henriette Bast a,b,c , Rafael C. Schick a,b, c , Thomas Koehler d,e , Franz Pfeiffer a,b,c,e a Chair of Biomedical Physics, Department of Physics, School of Natural Sciences, Technical University of Munich, 85748 Garching, Germany b Munich Institute of Biomedical Engineering, 85748 Garching, Germany c Institute for Diagnostic and Interventional Radiology, School of Medicine and Health, TUM Klinikum, Technical University of Munich (TUM), 81675 Munich d Philips Innovative Technologies, 22335 Hamburg, Germany e Munich Institute for Advanced Study, Technical University of Munich, 85748 Garching, Germany Corresponding e-mail address: [email protected] Corresponding author address: Henriette Bast, Chair of Biomedical Physics, Department of Physics, School of Natural Sciences, Technical University of Munich, 85748 Garching, Germany Keywords: X-ray imaging, Dark-field radiography, Low-dose simulation, radiation dose reduction 1 Introduction Grating-based X-ray dark-field radiography has shown significant potential in human chest imaging for the diagnosis of pulmonary diseases [ 2 , 3 , 4 , 5 ] . The dark-field image arises from small angle scattering of X-rays as they pass through microstructures such as pulmonary alveoli [ 6 ] , whereas the attenuation image is produced predominately by the photoelectric absorption and Compton scattering of the X-rays. A clinical prototype system for grating-based X-ray dark-field imaging has been developed at our institution and located at the university hospital Rechts der Isar, Munich, Germany. Since clinical dark-field radiography is a novel imaging procedure, the optimal balance between radiation dose and image quality has yet to be established. In accordance with the ALARA principle (As Low As Reasonably Achievable), simulations were initially performed to assess the impact of reduced radiation doses on image quality in dark-field radiographs, thereby paving the way for actual dose reduction in human dark-field imaging. Such simulated dose reduction has already been proven to enable dark-field imaging for the diagnosis of the Coronavirus SARS-CoV-2 (COVID-19) pneumonia with a 75\text{\,}\% -reduced median radiation dose of 10.5 µSv, as reported in our earlier study [ 1 ] . One key conclusion from that work was that further dose reduction is feasible, as algorithm-related artifacts became increasingly prominent at dose reductions of 75\text{\,}\% or more, significantly reducing image quality. In this study, we build on those findings by directly comparing simulated low-dose radiographs with corresponding actual low-dose measurements, acquired using a phantom, thus validating the reported low-dose simulation algorithm. 2 Materials and Methods The clinical prototype dark-field system located at the Technical University of Munich university hospital Rechts der Isar, Germany was used to to acquire attenuation and dark-field radiographs of the human thorax. The prototype’s main components are three optical gratings placed between an X-ray tube (MRC 200 0508 ROT-GS 1003; Philips Medical Systems, Germany) operated at 70\text{\,}\mathrm{k}\mathrm{V}\mathrm{p} and a flat-panel detector (PIXIUM 4343 F4; Trixell, France). 2.1 Image Acquisition and Reconstruction with the Clinical Dark-field Scanning System The clinical prototype system is a scanning system, that covers the full field-of-view (42 cm×42 cm) during image acquisition through a vertical scan of 6.5 s duration with an illuminated grating are of 42 cm×6.5 cm [ 2 , 4 ] . During this scanning procedure, a series of 195 exposures are measured by the detector at a frame rate of 30 Hz. After image reconstruction using a phase retrieval algorithm, both the simulated reduced dose images and the actual, experimental images were corrected for different sources of scatter [ 7 ] , dark-field bias, interferometer vibrations [ 8 ] , and beam hardening. Furthermore, the attenuation images were contrast enhanced and the dark-field images were denoised. Areas of direct x-ray exposure, surrounding the LUNGMAN, were concealed due to artifacts arising from detector non-linearity. The dark-field signal of the lung area, which was manually segmented based on the attenuation image, was averaged for the measured and simulated images. 2.2 Simulated Dose Reduction Algorithm The dose reduction algorithm is a modification of the motion artifact correction algorithm originally developed for the clinical dark-field system to reduce or eliminate motion artifacts (e.g. from the heartbeat) [ 9 ] . In summary, this algorithm partitions the acquired data into multiple virtual scans, each simulating a reduced height of the grating area, which effectively shortens the illumination time and thereby mitigating motion-related artifacts. Through a chi-squared analysis the most suitable virtual scan that minimizes motion in each pixel is chosen for motion correction. However, this virtual reduction in grating height, as illustrated in Fig. 1 , also corresponds to a simulated decrease in radiation exposure to both the detector and the patient, proportional to the extent of the height reduction. To enable simulated dose reduction, the algorithm was modified accordingly: The height reduction of the grating area is no longer fixed but variable, allowing the simulation of different levels of dose reduction. Additionally, instead of generating multiple virtual scans for the mitigation of motion artifacts, a single virtual scan is sufficient for the reconstruction of dark-field and attenuation radiographs with a reduced dose. Figure 1: Simulated Dose Reduction Algorithm. The dose reduction algorithm partitions the image data by virtually reducing the height of the original grating area (blue box) to 50\text{\,}\% (red box) or to 25 % (orange box), corresponding to dose reductions to 50 % or to 25 %, respectively. 2.3 Phantom Measurements To validate the dose reduction algorithm, measurements of the anthropomorphic chest phantom LUNGMAN (Kyoto Kagaku, Japan), filled with the accompanying urethane lungs, positioned the same way as patients, were taken. The two optional chest plates representing human fat tissue were attached to the phantom, thereby modeling a male person with a BMI of 29\text{\,}\mathrm{kg} . The phantom was imaged at 456\text{\,}\mathrm{m}\mathrm{A} , corresponding to the average radiation dose

Read the complete text