CRYSP: a Total-Body PET based on cryogenic cesium iodide crystals

S. R. Soleti, P. Dietz, R. Esteve, J. García-Barrena, V. Herrero, F. Lopez, F. Monrabal, L. Navarro-Cozcolluela, E. Oblak, J. Pelegrín, J. Renner, J. Toledo, S. Torelli, J. J. Gómez-Cadenas · 2025

Read CRYSP: a Total-Body PET based on cryogenic cesium iodide crystals by S. R. Soleti, P. Dietz, R. Esteve, J. García-Barrena, V. Herrero, F. Lopez, F. Monrabal, L. Navarro-Cozcolluela, E. Oblak, J. Pelegrín, J. Renner, J. Toledo, S. Torelli, J. J. Gómez-Cadenas (2025) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/arxiv-2507-09765.txt.

Introduction

Soleti and a large team proposed in 2025 a total-body medical scanner built from cooled cesium iodide crystals instead of pricey rare-earth materials. Total-body PET scanners are extremely sensitive, cutting scan time and radiation dose. Cost keeps them rare. Pure cesium iodide glows brightly when chilled near 100 kelvin, offering a cheaper detector material.

Tests showed energy resolution under seven percent and nanosecond-scale timing despite the crystal's slow glow. Simulations suggest millimeter-scale sharpness in all directions, including depth inside the crystal. The authors argue such a scanner could match high-end performance at far lower detector cost. That would widen access to total-body imaging in clinics and research centers alike.

Catalog record

CREATORS. R. Soleti, P. Dietz, R. Esteve, J. García-Barrena, V. Herrero, F. Lopez, F. Monrabal, L. Navarro-Cozcolluela, E. Oblak, J. Pelegrín, J. Renner, J. Toledo, S. Torelli, J. J. Gómez-Cadenas
YEAR2025
COLLECTIONPAPERS
FORMATTXT
FILE SIZE49 KB
VERSION1.0
LICENSEhttp://creativecommons.org/licenses/by/4.0/
RIGHTSCC BY
SOURCEarXiv

Source: https://arxiv.org/abs/2507-09765

Browse the full library · About the Tower of Babel

Full text

The opening of the work, from the library's hosted text:

CRYSP: a Total-Body PET based on cryogenic cesium iodide crystals S. R. Soleti, P. Dietz, R. Esteve, J. García-Barrena, V. Herrero, F. Lopez, F. Monrabal, L. Navarro-Cozcolluela, E. Oblak, J. Pelegrín, J. Renner, J. Toledo, S. Torelli, J. J. Gómez-Cadenas Abstract Total Body PET (TBPET) scanners have recently demonstrated the ability to significantly reduce both acquisition time and the administered radioactive dose, thanks to their increased sensitivity. However, their widespread adoption is limited by the high costs associated with the current available systems. In this context, pure cesium iodide (CsI) monolithic crystals, given their much lower cost compared with currently used rare-earth crystals, offer a promising solution to improve accessibility. When operated at cryogenic temperatures (approximately 100 K), CsI crystals exhibit one of the highest light outputs among inorganic crystals, around 10 5 10^{5} photons/MeV. This results in energy resolution below 7%, millimeter-scale spatial resolution (including accurate depth-of-interaction determination), and coincidence time resolution at the nanosecond level, despite their relatively slow scintillation decay time. This study demonstrates that a TBPET scanner based on cryogenic CsI monolithic crystals might have the potential to deliver high-performance imaging at a reduced cost compared with conventional systems, paving the way for broader deployment and accessibility. I Introduction Positron emission tomography (PET) is a powerful imaging technique central to medical diagnostic applications, employing positron-emitting radionuclides to track biologically active molecules. The annihilation of the positrons produced by the radiotracer with the electrons of the surrounding environment generates back-to-back gamma rays of 511 keV, which are detected to create detailed images of functional processes within the body. Fig. 1: Increase in component costs for a 70-cm, 100-cm, 140-cm and 200-cm system versus a system with 20 cm axial length; the y-axis is the system cost in relative units. Data taken from ref.. [ 1 ] . The last decade has witnessed the development of Total Body PET (TBPET) scanners with axial fields-of-view (AFOV) between 3 and 7 times that of conventional PET (CPET) apparatus. Such devices detect a large fraction of the emitted photons, thus dramatically increasing sensitivity, and allowing for simultaneous dynamic acquisition from all tissues of interest [ 2 , 1 ] . TBPETs also offer the flexibility of optimizing signal-to-noise ratio (SNR), reducing the acquisition time, or decreasing the amount of administered radioactive dose. Prime examples of such TBPETs are the United Imaging uEXPLORER [ 3 ] , and Siemens Biograph Vision Quadra [ 4 ] . The AFOV of uEXPLORER is 194 cm, while the AFOV of the smaller Quadra is 106 cm, to be compared with an AFOV of 20-30 cm typical of a CPET (e.g., the new generation General Electric Omni Legend 32 [ 5 ] has an AFOV of 32 cm). Both TPBET systems use arrays of pixelated lutetium-yttrium oxyorthosilicate (LYSO) crystals, read out by SiPMs and custom electronics, with very high sensitivity and excellent time-of-flight (TOF) resolution. The approximate increase in cost of the uEXPLORER (Quadra) compared with a typical CPET is roughly 8 (4) times (see fig. 1 ), well above what can be afforded by most hospitals. Fig. 2: The geometrical acceptance for a point-like source placed at the axial center, a point-like source placed 10 cm from one end of the scanner, and line sources of 10 cm, 100 cm and 200 cm length, as a function of the axial length. It is calculated in the transverse centre of a PET scanner with a diameter of 80 cm. The bulk of the cost in both systems comes from the scintillator, followed by the cost of the SiPMs and electronics. Of the three items, the cost of SiPMs, in large quantities, is roughly proportional to the total surface of chips deployed (not to the size of the chip), and thus proportional to the size and coverage of the scanner. However, the cost of SiPMs decreases about 5-10 % annually, with current prices about one third of those ten years ago. Thus their relative contribution to the total cost of future TPBETs will steadily decrease with time. Instead, the cost of crystals and that of the readout electronics has remained stable, and in the case of LYSO, significantly expensive. Thus, the first priority of any putative low-cost TBPET is to find significantly cheaper alternatives to LYSO. Notice that, in addition to the high cost, LYSO is a rare earth whose production and manufacturing is heavily concentrated. Last but not least, the cost of the readout electronics can be decreased for large scale production developing specific ASICs, and, importantly, reducing the number of electronics channels. Naturally, one also desires performance comparable to that of current PET scanners. This, in turn, requires quantifying the contribution of the extended axial length of a TBPET scanner to the quality of image reconstruction. Figure 2 shows the geometrical acceptance for two point-like sources (serving as rough approximations of small organs), one positioned at the center of a scanner with an 80 cm diameter and the other located 10 cm from one end, as a function of the axial length. For the centrally placed source, approximately 80% of the solid angle is already covered at a scanner length of 100 cm. In contrast, for the source near the edge (e.g., the brain in a TBPET), the acceptance quickly plateaus: photons emitted toward the open end of the scanner escape detection, and increasing the AFOV beyond this point offers no additional benefit. Overall, most of the gain in geometrical acceptance is achieved within the first 100 cm of axial coverage, with only marginal improvements beyond that length. The choice for a PET scanner of 100 cm length is also motivated by the approximate axial distribution of whole-body organs of interest (e.g., brain to pelvis) for typical human height. The figure shows that, with an extended source of 100 cm, one gains in geometrical acceptance from 50% (100 cm scanner), to about 80% (200 cm scanner), a modest increased compared with cost. Thus, an AFOV of about 100 cm as implemented by the Quadra appears as a good compromise between cost and performance and a reasonable target for any TPBET proposal. (a) (b) Fig. 3: (a) Compton scatter detection in PET. In Compton scattering a photon (in yellow) undergoes an interaction within the patient’s body where it changes direction and loses energy. Thus, the PET scanner reconstructs a wrong LOR (in red), affecting the signal-to-background ratio of the reconstructed image. Adapted from ref. [ 6 ] . (b) The residual energy E_{\gamma^{\prime}} of a 511-keV incoming photon after Compton scattering with a given angle \theta . Fig. 4: Mean parallax error \Delta\theta , defined as the angular distance between the ideal LOR and the reconstructed LOR, as a function of the LOR axial angle. The result was obtained by simulating a PET scanner with an axial length of 1 m and a diameter of 0.8 m, made of 3\times 3\times 22.4 mm 3 LYSO pixels. But is that extra length worth the added cost? Addressing this question requires consideration not only of the improved geometrical acceptance but also of the accompanying increase in gamma-ray absorption and Compton scattering (CS) within the patient’s body, as well as the effects of parallax error (PE). These effects become

Read the complete text