Development and validation of a high-fidelity full-spectrum Monte Carlo model for the Swiss airborne gamma-ray spectrometry system Journal: Nucl. Instrum. Methods Phys. Res.

David Breitenmoser, Alberto Stabilini, Malgorzata Magdalena Kasprzak, Sabine Mayer · 2025

Read Development and validation of a high-fidelity full-spectrum Monte Carlo model for the Swiss airborne gamma-ray spectrometry system Journal: Nucl. Instrum. Methods Phys. Res. by David Breitenmoser, Alberto Stabilini, Malgorzata Magdalena Kasprzak, Sabine Mayer (2025) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/arxiv-2502-02102.txt.

Introduction

Breitenmoser, Stabilini, Kasprzak, and Mayer build in twenty twenty-five a detailed computer model of Switzerland's airborne radiation survey system. Aircraft scanners map radioactive fallout fast after accidents, but calibrating them with real sources covers only a few isotopes, leaving most accident nuclides unmeasured. Their simulation models the aircraft itself in fine detail plus the detector's light response, predicting the full measured spectrum for any source.

Tests in near, middle, and far setups showed accuracy fit to replace the old empirical calibration. The model can mint signature spectra for any nuclide without handling strong sources, cutting cost, time, and waste. The authors pitch it as groundwork for richer airborne data methods, from emergency mapping to nuclear security.

Catalog record

CREATORDavid Breitenmoser, Alberto Stabilini, Malgorzata Magdalena Kasprzak, Sabine Mayer
YEAR2025
COLLECTIONPAPERS
FORMATTXT
FILE SIZE58 KB
VERSION1.0
LICENSEhttp://creativecommons.org/licenses/by/4.0/
RIGHTSCC BY
SOURCEarXiv

Source: https://arxiv.org/abs/2502-02102

Browse the full library · About the Tower of Babel

Full text

The opening of the work, from the library's hosted text:

Development and validation of a high-fidelity full-spectrum Monte Carlo model for the Swiss airborne gamma-ray spectrometry system Journal: Nucl. Instrum. Methods Phys. Res. David Breitenmoser, Alberto Stabilini, Malgorzata Magdalena Kasprzak, Sabine Mayer Abstract Airborne Gamma-Ray Spectrometry (AGRS) is a critical tool for radiological emergency response, enabling the rapid identification and quantification of hazardous terrestrial radionuclides over large areas. However, existing calibration methods are limited to only a few gamma-ray sources, excluding most radionuclides released in severe nuclear accidents and nuclear weapon detonations, which compromises effective response and accurate risk assessments in such incidents. Here, we present a high-fidelity Monte Carlo model that overcomes these limitations, offering full-spectrum calibration for any gamma-ray source. Unlike previous approaches, our model integrates a detailed mass model of the aircraft and a calibrated non-proportional scintillation model, enabling accurate event-by-event predictions of the spectrometer’s response to arbitrarily complex gamma-ray fields. Validation measurements in near-, mid-, and far-field scenarios demonstrate that the developed model not only effectively addresses the large deficiencies of previous approaches, but also achieves the accuracy and precision required to supersede traditional empirical calibration methods. These results mark a major advancement in AGRS. The developed calibration methodology not only allows for the generation of high-fidelity spectral signatures for any gamma-ray source, but also reduces calibration time and costs, minimizes reliance on high-intensity calibration sources, and eliminates the generation of related radioactive waste. The approach presented here serves as a critical step toward integrating advanced full-spectrum data reduction methods for AGRS, which could unlock promising new capabilities for these systems beyond emergency response, including the quantification of the cosmic-ray flux in the atmosphere for geophysical research and the identification of trace-level airborne radionuclides in nuclear security applications. 1 Introduction By mounting high-volume gamma-ray spectrometers in an aircraft, Airborne Gamma-Ray Spectrometry (AGRS) systems enable the localization, identification, and quantification of terrestrial gamma-ray sources over extensive areas \mathcal{O}($1\text{\times}{10}^{8}$)\,${\mathrm{m}}^{2}$ within \mathcal{O}($1\text{\times}{10}^{2}$)\,$\mathrm{min}$ [ 1 ] . This unique capability makes AGRS essential for a wide range of applications, including radiological emergency response, geological mapping, mineral exploration, and environmental monitoring [ 2 , 3 , 4 ] . The accuracy and sensitivity of AGRS systems rely heavily on the adopted calibration protocols. The International Atomic Energy Agency (IAEA) has standardized these methods in two technical reports [ 5 , 6 ] , which utilize simplified physics models combined with empirical calibration. These methods, while useful, have critical limitations. Current empirical calibration methods are restricted to a limited range of radionuclide sources and source-detector configurations. Comprehensive calibration can require up to 5\text{\,}\mathrm{d} to cover all relevant source-detector configurations and often involves high costs for source preparation and/or decommissioning [ 7 , 8 , 9 ] . Additionally, these methods are subject to significant systematic uncertainties, primarily originating from varying radiation backgrounds and required analytical corrections. The physics models employed, on the other hand, are only reliable for photon energies \gtrsim\!$400\text{\,}\mathrm{keV}$ [ 5 , 6 ] . Moreover, these models are limited to narrow spectral bands, which greatly reduces the AGRS sensitivity by excluding much of the available information encoded in the pulse-height spectra [ 10 , 11 ] . These limitations in current AGRS calibration protocols significantly compromise the quality and reliability of the derived data for emergency response applications. This became particularly evident in the severe nuclear accident at the Fukushima Daiichi nuclear power plant in 2011, where the current calibration limitations in AGRS severely hindered effective response and accurate risk assessment [ 12 ] . Improving the current calibration and data evaluation protocols in AGRS to address the discussed limitations in the capabilities of AGRS systems is therefore of immediate relevance for public safety. To overcome the discussed limitations in AGRS calibration and data evaluation, prior studies have proposed following a numerical instead of an empirical approach, specifically high-fidelity Monte Carlo simulations [ 13 , 14 ] . Monte Carlo models offer the potential to predict the full-spectrum gamma-ray spectrometer response without restrictions in the source type or source-detector geometry [ 15 , 16 , 17 , 18 ] . However, implementing Monte Carlo based calibration and data evaluation for AGRS has proven challenging due to the model’s complexity and high computational cost [ 13 , 14 , 19 , 20 , 21 , 22 , 23 ] . These difficulties arise mainly from the large simulation domain \mathcal{O}($1\text{\times}{10}^{10}$)\,${\mathrm{m}}^{3}$ and the modeling effort required to develop a sufficiently accurate mass model of the aircraft platform with characteristic lengths of \mathcal{O}($1\text{\times}{10}^{1}$)\,$\mathrm{m}$ . To reduce model complexity and computational costs, earlier studies either omitted the aircraft entirely or represented it as primitive solids in their Monte Carlo models [ 13 , 14 , 19 , 23 , 20 , 21 , 22 ] . These simplifications have proven to introduce significant systematic errors in the simulations, in particular at photon energies \lesssim\!$1\text{\times}{10}^{2}\text{\,}\mathrm{keV}$ with relative errors exceeding 200\text{\,}\mathrm{\%} [ 13 , 22 ] . Moreover, most studies followed traditional IAEA methods, focusing on narrow spectral bands through a hybrid approach that integrates monoenergetic flux models with Monte Carlo simulations, rather than predicting the full-spectrum response of the gamma-ray spectrometer [ 14 , 19 , 23 , 20 , 21 ] . As a result, Monte Carlo based methods have not yet reached the accuracy or sensitivity required to supersede traditional AGRS methodologies, leaving the fundamental limitations in AGRS discussed above unresolved. The main scope of this work is the development and validation of an advanced full-spectrum Monte Carlo model for the Swiss AGRS system, aiming to overcome the fundamental limitations of traditional AGRS calibration methods, particularly their restriction to a limited range of radionuclide sources, source-detector configurations, and gamma-ray energies. The developed Monte Carlo model features two key innovations: (1) a high-fidelity mass model of the entire aircraft to reduce systematic errors noted in previous studies, and (2) the integration of an advanced non-proportional scintillation model (NPSM) that enables accurate event-by-event prediction of the full-spectrum response of inorganic scintillators. Together, these advancements allow us to perform full-spectrum calibration of the Swiss AGRS system for both point and extended gamma-ray sources with an unprecedented level of accuracy. We demonstrate this through a series of validation measurements conducted under various field conditions with both natural and man-made radionuclides. The validation of the gamma-ray spectrometer mass model under laboratory conditions has been presented in earlier studies [ 24 , 25 ] . 2 Methods 2.1 Swiss AGRS system Here, we review the technical specifications of Switzerland’s currently operational AGRS system, Radiometrie Land-Luft (RLL), hereafter referred to as the Swiss AGRS system. A comprehensive discussion of its development, operational capabilities, and integration within Swiss civil and military organizations is provided by Breitenmoser 2024 . Figure 1: Overview of the Swiss AGRS system. This illustration highlights a close-up view of the spectrometer location within the fuselage of the AS332M1 helicopter, the detector box, the scintillation detectors as well as the support systems. In addition, selected secondary helicopter sensors accessible to the AGRS system, i.e. global navigation satellite system (GNSS) receiver and radar altimeter, as well as the universal mobile telecommunications system (UMTS) antenna are marked. The line drawing of the AS332M1 helicopter was adapted from Jetijones, CC BY 3.0 , via Wikimedia Commons. As illustrated in Fig.

Read the complete text