Influence of color correction on pathology detection in Capsule Endoscopy
Bidossessi Emmanuel Agossou, Marius Pedersen, Kiran Raja, Anuja Vats, Pål Anders Floor · 2025
Read Influence of color correction on pathology detection in Capsule Endoscopy by Bidossessi Emmanuel Agossou, Marius Pedersen, Kiran Raja, Anuja Vats, Pål Anders Floor (2025) in the Tower of Babel, a free online library of public-domain works preserved in clean, searchable text. No sign-up. The full text is served as a plain-text file at a predictable URL, where text is present: https://antarctic-labs.com/tower-of-babel/arxiv-2502-00076.txt.
Introduction
Bidossessi Agossou and co-authors test how color fixes in capsule camera images affect AI disease spotting, in a 2025 paper. Patients swallow a tiny camera that films the gut. Deep learning models scan that film for disease. Color quality in training data can sway those models.
The team made two color-corrected versions of a known data set. They trained two detection models on the raw and fixed versions. Color fixes led the models to draw larger boxes that overlapped the true marks more. For some diseases, fixes also raised false alarms. The mixed result warns that image cleanup is not free of cost. Clinics should test such steps before adopting them in screening tools.
Catalog record
| CREATOR | Bidossessi Emmanuel Agossou, Marius Pedersen, Kiran Raja, Anuja Vats, Pål Anders Floor |
|---|---|
| YEAR | 2025 |
| COLLECTION | PAPERS |
| FORMAT | TXT |
| FILE SIZE | 25 KB |
| VERSION | 1.0 |
| LICENSE | http://creativecommons.org/licenses/by/4.0/ |
| RIGHTS | CC BY |
| SOURCE | arXiv |
Source: https://arxiv.org/abs/2502-00076
Browse the full library · About the Tower of Babel
Full text
The opening of the work, from the library's hosted text:
Influence of color correction on pathology detection in Capsule Endoscopy Bidossessi Emmanuel Agossou, Marius Pedersen, Kiran Raja, Anuja Vats, Pål Anders Floor Abstract Pathology detection in Wireless Capsule Endoscopy (WCE) using deep learning has been explored in the recent past. However, deep learning models can be influenced by the color quality of the dataset used to train them, impacting detection, segmentation and classification tasks. In this work, we evaluate the impact of color correction on pathology detection using two prominent object detection models: Retinanet and YOLOv5. We first generate two color corrected versions of a popular WCE dataset (i.e., SEE-AI dataset) using two different color correction functions. We then evaluate the performance of the Retinanet and YOLOv5 on the original and color corrected versions of the dataset. The results reveal that color correction makes the models generate larger bounding boxes and larger intersection areas with the ground truth annotations. Furthermore, color correction leads to an increased number of false positives for certain pathologies. However, these effects do not translate into a consistent improvement in performance metrics such as F1-scores, IoU, and AP50. The code is available at https://github.com/agossouema2011/WCE2024 . 1 Introduction Digestive system diseases are widespread around the world and cause considerable distress which can be fatal. In 2019, about 2276.27 million estimated prevalent cases and 2.56 million deaths were counted for digestive system diseases [ 1 ] . Endoscopy is the procedure commonly used to visualize the gastrointestinal tract and detect diseases [ 2 ] . However, traditional endoscopy is uncomfortable and painful for patients [ 3 ] , discouraging it for wider preventive screening programs. A new type of gastrointestinal endoscopy, known as Wireless Capsule Endoscopy (WCE) [ 4 ] , was introduced in 2000, which is an alternative and minimally invasive screening method. In WCE, the patient swallows a pillcam (pill-sized camera) which moves through the gastrointestinal tract and records videos [ 5 ] . The whole process can take on average eight to twelve hours, generating around 60,000 frames [ 6 ] . Considering the long procedure duration, the demand placed on gastroenterologists for WCE screening is huge, leaving room for missing pathologies. Therefore, methods to improve pathology detection accuracy and reduce the time spent in screening WCE images are of great importance. However, WCE has some drawbacks compared to traditional endoscopy like low frame rate and images of low resolution with poor quality [ 7 ] . In addition, colors are distorted in WCE, making it challenging to perform color reproduction and color consistency [ 8 ] . Correcting or enhancing the color of the WCE images before training a deep learning model could help improve pathology detection accuracy. Faithfulness to color of human tissues and blood vessels is key to accurately detecting abnormalities that often deviate from normal in their appearances [ 8 ] . Post-processing algorithms can distort colors that may lead to misinterpretation. Thus making color reproduction accuracy and consistency all the more significant for human inspection as well as automated diagnosis. Color accuracy refers to the ability of a system to produce exact color matches from input to output, while color consistency refers to the ability to produce image data with a similar response as that in a human interpreter [ 9 ] [ 8 ] . Maintaining color accuracy through color correction along with resolution and contrast are important for diagnostic use [ 9 ] . This work evaluates the influence of color correction on pathology detection in WCE. Detection in our context means pathology localization inside the image using bounding boxes. To the best of our knowledge, little attention is given to this direction in the literature. The contributions of this work are the following: • We present two new color corrected versions of the publicly available SEE-AI [ 10 ] dataset. The dataset will be made publicly available to other researchers for investigating approaches for classification and detection tasks. • We present a detailed analysis evaluating how slight changes in colors could potentially impact the behaviour of detection models in WCE. The analysis points to strengths and weakness of color correction for WCE for future works to build upon. 2 Related works Object detection has gained huge attention [ 11 ] and consists of identifying objects or instances of objects in an image or a video using bounding boxes. There are two main groups of object detection methods: one-stage detectors and two-stage detectors [ 12 ] [ 13 ] . One-stage objects detectors train models using a large amount of ground truth data. i.e., accurate human-annotated data. They require only a single pass through the neural network and predict all the bounding boxes in one go. One-stage methods include You Only Look Once (YOLO) [ 14 ] , and Single Shot Detector (SSD) [ 15 ] . The two-stage detectors on the other hand generate a set of regions of interest with a region proposal network, and then perform classification and bounding box regression. Popular two-stage models include Faster-RCNN [ 16 ] and Mask R-CNN [ 17 ] . Ding et al. [ 18 ] used a Convolutional Neural Network (CNN) based model to first classify WCE images into normal and abnormal frames, and further categorize the abnormal images into different pathology classes including inflammation, ulcer, polyps, lymphangiectasia, and bleeding. Aoki et al. [ 19 ] used a CNN-based model to detect blood content in WCE images. A CNN-based detection system for various abnormalities using WCE videos from multiple centers was developed by Aoki et al. [ 20 ] . The system combines four CNN-based models: three SSD models and one ResNet50, to detect among other, mucosal breaks, angioectasia, polyps, submucosal tumors, venous structures, blood contents. Yokote et al. [ 10 ] created an annotated dataset for WCE containing 12 types of small bowel pathologies. In addition, an object detection model based on YOLOv5 was used to detect the pathologies. Cherepkova and Hardeberg [ 21 ] investigated the impact of image color enhancement on the performance of a lesion classification algorithm for skin cancer detection. Six different image enhancement algorithms were selected to investigate which algorithms have more influence on lesion detection improvement, focusing especially on color correction. A CNN based classifier and support vector machine were used on a set of dermoscopy image dataset. Color enhancement generally yielded slightly better results for color versions of the images in their analysis [ 21 ] . In order to solve color distortion problem in WCE, Watine et al. [ 8 ] considered the color correction of endoscopy videos through post-processing using a dedicated color checker to select the colors that typically appear in the human colon. The developed color checker based on the colon gamut was named ColonColorChecker (CCC). Two color correction matrices were derived, one based on a standard Gretag Macbeth ColorChecker (CC) and the other bases on the CCC. These were applied to enhance color in WCE images. From their objective and subjective evaluations with two experienced gastroenterologists, the color corrections provide a clear improvement of color in clinical WCE videos. However, the color corrections decreased the contrast in the color corrected videos reducing the diagnostic value of the approach [ 8 ] . 3 Methodology Motivated