High-precision beam profile measurement with a microchannel-plate detector in the high magnetic field of the WISArD experiment S. Lecanuet, X. Fléchard, P. Alfaurt, P. Ascher, D. Atanasov, B. Blank, L. Daudin, H. DePreaumont, M. Gerbaux, J. Giovinazzo, S. Grévy, G. Guignard, J. Ha, C. Knapen, S. Lechner, A. Lépine, J. Lory, J. Perronnel, M. Pomorski, M. Roche, C. Roumegou, N. Severijns, Y. Son, S. Vanlangendonck, M. Versteegen, D. Zakoucky Abstract We present the development and characterization of a compact low-energy ion beam diagnostic for the WISArD (Weak Interaction Studies with Ar 32 \mathrm{{}^{32}Ar} Decay) experiment at ISOLDE/CERN. The microchannel plate (MCP) detector, which is configured in a Z-stack and has a resistive position sensitive anode, was tested with both stable and radioactive beams. This work focuses on the image reconstruction method, which corrects the pincushion distortion inherent to the square-shaped resistive anode, and investigates the influence of the magnetic field on the detector performance. Our results demonstrate that the detector achieves beam profile measurements with sub-millimeter accuracy, while coping with the spatial and high-magnetic field (4 T) constraints of the experiment. These capabilities meet the precision requirements of the WISArD experiment for extracting the modified beta-neutrino angular correlation coefficient, a ~ β ​ ν \tilde{a}_{\beta\nu} , with an uncertainty of 1‰. 1 Introduction Microchannel plate (MCP) detectors are widely used in atomic, nuclear, high-energy and anti-matter physics to detect single particles such as photons, electrons, and ions. When coupled to a position sensitive anode, these detectors provide both the detection time and the spatial coordinates of particles, achieving typical accuracies better than 100 ps and 100 µm for time and position, with detection efficiencies larger than 50% for ions [ 14 ] . Such capabilities make MCP detectors particularly suitable for precise measurements of the full momentum of low-energy particles [ 18 , 15 ] and for beam imaging applications [ 4 , 3 , 19 ] . In the WISArD (Weak Interaction Studies with \mathrm{{}^{32}Ar} Decay) experiment at ISOLDE/CERN, which probes the electroweak interaction through high-precision measurements in nuclear beta decay, knowledge of the radioactive beam profile in the experimental setup is crucial. The goal of the experiment is the determination of the modified beta-neutrino angular correlation coefficient, \tilde{a}_{\beta\nu} [ 9 ] , in the beta decay of \mathrm{{}^{32}Ar} , with a precision of 1‰ in order to probe exotic current beyond the Standard Model [ 5 ] . It relies on the measurement of the energy shift of the delayed protons emitted after beta decay, as a function of the relative angle between these protons and the beta particles [ 1 ] . In the Monte-Carlo simulations used to extract \tilde{a}_{\beta\nu} , this relative angle critically depends on the location of the decay source. The ISOLDE facility delivers a 30 keV radioactive ion beam of \mathrm{{}^{32}Ar} , which is implanted into a thin aluminized mylar catcher foil at the center of the detection system. The analysis of the WISArD proof-of-principle experiment revealed a systematic error of \Delta\tilde{a}_{\beta\nu}=4 ‰ resulting from the 3 mm uncertainty estimated on both the position and the radius of the implantation region [ 1 ] . To address this significant source of systematic error, a high-precision beam diagnostic with sub-millimeter accuracy was developed for integration into the upgraded experimental setup [ 2 ] . The WISArD detection tower, shown in Fig. 1 , is inserted in the former WITCH [ 16 ] superconducting magnet which provides the high magnetic field (4 T) required for the experiment but offers limited space for such a diagnostic. We refer the reader to Ref. [ 2 ] for a detailed description of the WISArD detection setup. Given the above mentioned constraints, a compact MCP detector coupled to a square resistive anode was built and tested. This article describes the detector design (Sec. 2 ), details the image reconstruction and the detector’s performance (Sec. 3 ) Finally, Sec. 4 presents the beam profile measurements obtained during the 2025 run at ISOLDE with radioactive \mathrm{{}^{32}Ar} ions, and discuss their impact on the extraction of the beta-neutrino angular correlation coefficient \tilde{a}_{\beta\nu} . Figure 1: CAD view of the WISArD detection setup with the radioactive beam in fluorescent green implanted on the catcher. The MCP has been integrated at the bottom of the tower (here in out position). See Ref. [ 2 ] for further details. 2 MCP detector design 2.1 Choice of the position sensitive anode and MCP assembly The design of the beam monitor is constrained by two main factors: the high magnetic field and the limited space within the WISArD tower. An MCP-based detector was selected as it is well-suited for detecting ions with typical incident energies of several keV, while their compact size is advantageous for integration into the WISArD setup. However, it is well established that strong magnetic fields significantly reduce the gain of MCPs. This effect is due to the shortening of electron trajectories between two impacts on the microchannel surface leading to the reduction of secondary emission within the microchannels [ 13 ] . Typically, a magnetic field of the order of 1 T can reduce the gain of standard 25 µm channel MCPs by one order of magnitude. By using a smaller channel diameter and bias angle [ 12 ] one can mitigate significant gain loss from operating the detector in an environment of 4 T (targeted field for the online measurements). As a compromise between costs and performances, the F1551-01 from Hamamatsu, with a 17.9 mm plate diameter, 12 µm channel diameter and 8° bias angle was chosen. Three MCPs were mounted in a Z-stack configuration, which provides higher gain than the standard chevron configuration and increases the total charge emitted by the MCP assembly. For position-sensitive readout, several anode types are commonly used with MCPs, including: delay lines, phosphor screens, backgammon anodes, and resistive anodes. The strong magnetic field of WISArD confines the electron cloud emitted by the back MCP to the scale of the microchannels, making anodes requiring significant electron cloud spreading (such as backgammon or delay lines) unsuitable. Phosphor screens, which require a CCD camera and optical coupling, were incompatible with the available space. Resistive anodes [ 8 , 11 ] are compact and compatible with a high magnetic field but commercially available detectors based on the Gear-anode geometry [ 8 ] were too large for the WISArD tower with a maximum diameter of 12 cm. Instead, a custom-made square-shaped resistive anode was fabricated. The resistive layer was created by mixing graphite powder with glyceride-based paint at a mass ratio of 1.00 (paint) to 1.33 (graphite), yielding a homogeneous mixture with suitable viscosity. This mixture was applied to a 200 µm deep, 16 mm × 16 mm square pocket milled into a PEEK 1 1 1 Polyether ether ketone substrate, which serves as the anode holder. Excess paint was removed with a flat ruler to ensure a uniform 200 µm thickness. After drying, the per-square resistance R □ of the anodes that were produced typically ranged between 1 kΩ and 2 kΩ. 2.2 Mechanical design The MCP detector was designed to be inserted upstream of the detector tower (Fig. 1 ). It is mounted on one of the vertical rotatable rods, allowing it to be moved out of the central axis position to a side position, where it does not intercept the incoming beam or the decay products emitted from the catcher, as shown in Fig. 2 . These constraints limited the detector assembly to a maximum thickness of 15 mm and a maximum width of 25 mm. The resulting compact design is shown in Fig. 3 . Note that in Fig. 2 and Fig. 4 the calibration mask is the 2024 geometry having lines merging in the center, while the data uses in 2025 version with a central hole in the center. (a) (b) Figure 2: Bottom view of the WISArD experimental tower with the MCP position out (a) and in (b). The catcher foil stopping the beam during data taking is held by the white PEEK ring at the center of the tower. The calibration mask geometry is the 2024 version. Figure 3: Exploded view of the MCP detector assembly. See text for details. The main holder that houses the resistive anode as well as the MCP centering rings are made of PEEK. Passive electronic components (surface-mount devices) required for fast signal decoupling and detector operation are directly mounted on two printed circuit boards (PCBs) fixed to the main PEEK holder (see Fig. 4 ). The top PCB is used for the front MCP bias voltage and to collect the back MCP fast charge signal, while the bottom PCB supplies the acceleration voltage for the anode and collects charge signals from its four corners. Connection to the four corners is ensured by four thin stainless-steel strips pressed against the anode surface. A 0.5 mm thick aluminum mask with a 2 mm pitch pattern (1.4 \times 1.4 mm 2 holes) is fixed in front of the first MCP for spatial calibration. As in Ref. [ 10 ] , the mask remains permanently mounted on the detector to provide reliable online calibration, even in the event of small electronic gain drifts. With only 36 % transparency, the mask can also function as a Faraday cup when tuning a stable beam with intensities in the pA range. Figure 4: Picture of the front (a) and back (b) side of the detector with the 2024 mask geoemtry. 2.3 Detector read-out Figure 5: MCP detector electronic read-out, see text for details. The circuit diagram of the MCP detector is shown in Fig. 5 . Bias voltages are supplied to the MCPs and the anode through 510 kΩ resistors. Typical operating voltages for the front MCP range from - 2.0 kV to - 3.2 kV (HVMCP in Fig. 5 ), while the back MCP is connected to ground and the anode is supplied with +100 V (HVAnode in Fig. 5 ). Fast signals from the back MCP and from the four anode corners are decoupled from their DC voltage using a 2.2 nF capacitor and a 1 MΩ resistor connected to ground. The back MCP charge signal (SMCP in Fig. 5 ) and the anode signals (SAx in Fig. 5 ) are then sent through 50 Ω coaxial cables to a x10 gain fast current preamplifier and four charge preamplifiers, respectively. The charge preamplifiers, with a gain of \sim 1 V/pC, are tuned for short signal decay times ( < 1 µs) to accommodate high counting rates. The output signal of the MCP fast preamplifier is digitized and integrated using a CARAS board from the FASTER data acquisition system (DAQ) [ 6 ] . The four anode signals from the charge preamplifiers are digitized and numerically filtered using a MOSAHR board from the same FASTER DAQ. All data was individually time-stamped for both online and offline trigger analysis with a time resolution of 8 ns. 3 Characterization with stable beam The MCP detector was characterized using the offline ion source of WISArD, which provides a stable 30 keV \mathrm{{}^{39}K} + ion beam with intensities up to a few nA. The beamline optics were optimized using a low-intensity beam (a few pA) and a Faraday cup located at the end of the vertical WISArD beamline, downstream of the detection tower (for details see Ref. [ 2 ] ). To prevent detector saturation and potential damage the beam intensity was reduced using a set of three 90% attenuation grids installed at the entrance of the vertical beamline, resulting in a detector counting rate of a few 10^{3} counts per second (cps). This attenuation minimized pile-up effects and ensured safe operation. The beam was then scanned across the detector surface using x- and y-steerers, controlled by a dedicated PyEpics [ 17 ] program to map the full image of the calibration mask. Detector testing began with a front MCP bias voltage of - 2.0 kV in the absence of a magnetic field (Sec. 3.1 to 3.3 ), followed by a performance evaluation with magnetic fields ranging from 1 T to 4 T (Sec. 3.4 ). 3.1 Event selection Figure. 6 shows the time difference histograms between each of the four corner signals of the anode and the MCP signal used as a trigger. The main coicidence peak is located at - 170 ns. The plateau before the main coincidence peak, is caused by random coincidences. Each channel has its own intrinsic dead time, either imposed by the QDC time integration window (200 ns for the SMCP) or by the ADC signal sampling time (1080 ns for the SAx). The latter leads to strong suppression of random coincidences after the main peak until \Delta T=950 ns. Other peaks, occurring every 100 ns after the main one, arise from reflections in the cables. For event reconstruction, a time selection window [ - 190, - 100] ns is applied, indicated in Fig. 6 with vertical dashed lines. This time selection window removes events with badly reconstructed position, shown in Fig. 7 with red points. These represent \sim 8\% of the data. Figure 6: Time difference between the signals of each corner and the MCP signal. The limits [ - 190, - 100] ns of the time coincidence window used for event selection are indicated with red dashed vertical lines. 3.2 Image reconstruction and position calibration With resistive anodes using the Gear-geometry, the MCP detector image reconstruction can be easily obtained by comparing the charge collected at the four corners [ 11 ] : \displaystyle X_{d} \displaystyle= \displaystyle\frac{\displaystyle-C_{1}+C_{2}+C_{3}-C_{4}}{\displaystyle C_{\Sigma}} (1) \displaystyle Y_{d} \displaystyle= \displaystyle\frac{\displaystyle-C_{1}-C_{2}+C_{3}+C_{4}}{\displaystyle C_{\Sigma}} where C_{\Sigma} is the sum of the four charges. X_{d} and Y_{d} provide relative impact coordinates associated with the four collected charges C_{i} . Figure 7: MCP detector scan image with stable beam reconstructed using Eq. ( 1 ). Red points represent events outside the coincidence time window (see Sec. 3.1 ). Figure. 7 shows the reconstructed image using Eq. ( 1 ) from a scan of the MCP detector surface using the stable ion beam. The major pincushion distortion originates from using a square-shaped resistive anode [ 7 ] . The image is, moreover, off-centered and the bottom right corner appears quenched. This is caused by gain differences between the four corners, particularly with the bottom right corner, which has a higher gain. This distortion can be corrected using the logarithm of the normalized corner pulse heights. Keeping the same formalism as in Eq. ( 1 ), the corrected position (X,Y) can then be expressed as: \displaystyle X \displaystyle= \displaystyle-\frac{\displaystyle-c_{1}+c_{2}+c_{3}-c_{4}}{\displaystyle c_{1}+c_{2}+c_{3}+c_{4}} (2) \displaystyle Y \displaystyle= \displaystyle-\frac{\displaystyle-c_{1}-c_{2}+c_{3}+c_{4}}{\displaystyle c_{1}+c_{2}+c_{3}+c_{4}} where c_{i}=\ln(C_{i}/C_{\Sigma}) . Figure 8: MCP detector scan image with the stable ion beam reconstructed using Eq. ( 2 ) with the stable beam. The red contour is obtained for F_{scan}(X,Y)=A/2 (see text for details). Figure. 8 shows the result of the logarithmic correction using Eq. ( 2 ). The pincushion distortion is significantly reduced, but further corrections are required to achieve the desired precision on the ion hit position. The most straightforward reference points on the reconstructed image are the corner positions of each hole of the calibration mask. These reference points were used for the detector calibration. The coordinates (X_{jABCD},Y_{jABCD}) for each hole j in Fig. 8 were first determined manually. Then, a 2D function was built to determine these coordinates more precisely using a fit of the mask image. The contour of a deformed hole j was defined by four linear functions, f^{AB}_{j} , f^{BC}_{j} , f^{CD}_{j} and f^{DA}_{j} , derived from the coordinates of the four corners of the hole labeled A , B , C and D , as shown in Fig. 9 : \displaystyle f^{AB}_{j}(X) \displaystyle=\frac{\displaystyle Y_{jA}-Y_{jB}}{\displaystyle X_{jA}-X_{jB}}(X-X_{jA})+Y_{jA} (3) \displaystyle f^{BC}_{j}(Y) \displaystyle=\frac{\displaystyle X_{jB}-X_{jC}}{\displaystyle Y_{jB}-Y_{jC}}(Y-Y_{jB})+X_{jB} (4) \displaystyle f^{CD}_{j}(X) \displaystyle=\frac{\displaystyle Y_{jC}-Y_{jD}}{\displaystyle X_{jC}-X_{jD}}(X-X_{jC})+Y_{jC} (5) \displaystyle f^{DA}_{j}(Y) \displaystyle=\frac{\displaystyle X_{jD}-X_{jA}}{\displaystyle Y_{jD}-Y_{jA}}(Y-Y_{jD})+X_{jD}\,\,. (6) The mask transmission can then be expressed as: I_{j}(X,Y)=\begin{cases}1,\text{if }f^{AB}_{j}X\;\land\;f^{CD}_{j}>Y\;\land\;f^{DA}_{j}