Instrumentation and control in a laboratory-scale process for CO2 capture post-combustion Authors: María Fernanda Silva León, Erick Omar Castañeda Magadán, Miriam Navarrete Procopio, Víctor Manuel Zezatti Flores, Ángel Tlatelpa Becerro Published: 2026 ABSTRACT Graphical abstract Specifications table Hardware nameLow-cost DAQ for a post-combustion process.Subject area•Electrical − electronic engineering•System data acquisition (DAQ)•Chemical engineering•Educational tools and open-source alternatives to existing infrastructureHardware type•Measuring process variables and in-lab sensors•Field measurements and sensors•Electrical engineering and computer scienceClosest commercial analog•Development boards (Arduino, Blue Pill, Raspberry Pi)•Programmable Logic Controller•National Instruments•Agilent Technology.Open source licenseCreative Commons Attribution 4.0 International (CC BY 4.0)Cost of hardwareUSD 355.5Source file repositoryZenodo [https://doi.org/10.5281/zenodo.20735751] Hardware nameLow-cost DAQ for a post-combustion process.Subject area•Electrical − electronic engineering•System data acquisition (DAQ)•Chemical engineering•Educational tools and open-source alternatives to existing infrastructureHardware type•Measuring process variables and in-lab sensors•Field measurements and sensors•Electrical engineering and computer scienceClosest commercial analog•Development boards (Arduino, Blue Pill, Raspberry Pi)•Programmable Logic Controller•National Instruments•Agilent Technology.Open source licenseCreative Commons Attribution 4.0 International (CC BY 4.0)Cost of hardwareUSD 355.5Source file repositoryZenodo [https://doi.org/10.5281/zenodo.20735751] •Electrical − electronic engineering Electrical − electronic engineering •System data acquisition (DAQ) System data acquisition (DAQ) •Chemical engineering Chemical engineering •Educational tools and open-source alternatives to existing infrastructure Educational tools and open-source alternatives to existing infrastructure •Measuring process variables and in-lab sensors Measuring process variables and in-lab sensors •Field measurements and sensors Field measurements and sensors •Electrical engineering and computer science Electrical engineering and computer science •Development boards (Arduino, Blue Pill, Raspberry Pi) Development boards (Arduino, Blue Pill, Raspberry Pi) •Programmable Logic Controller Programmable Logic Controller •National Instruments National Instruments •Agilent Technology. Agilent Technology. Hardware in context Thermal power plants are the main emitters of CO2, with a significant carbon footprint estimated at 6.52 million tons. Of this large volume of emissions, 90.23% comes directly from the burning of fossil fuels such as coal, oil, and natural gas, contributing significantly to total greenhouse gas emissions. As a result, these emissions directly contribute to global climate change, which exacerbates phenomena such as rising Earth's average temperature and altered weather patterns [2], [3]. To mitigate these environmental impacts, considerable research has focused on developing carbon capture technologies to reduce CO2 emissions from industrial sources. A CO2-CP employs amine-based solvents to mitigate CO2 emissions from thermal power plants [4], [5], [6]. The core components of the post-combustion capture and regeneration process are packed columns. An absorption column with specific dimensions can be filled randomly; conversely, a packed desorption column is utilized to regenerate the absorbent solution [7]. Due to the efficiency of the absorption and regeneration processes strongly depends on stable operating conditions, accurate instrumentation and process control become essential. Chemical process instrumentation and control emphasize the automation of variables such as temperature, flow, and chemical composition to ensure process stability and safety. Likewise, in CO2-CP, instrumentation and control aim to optimize operational efficiency, manage changes in operating conditions, and lower operating costs through advanced control strategies [8], [9], [10], [11]. The implementation of temperature acquisition and control ensures stability within fixed limits, maintains consistent conditions for different processes, and improves efficiency. [12]. At the industrial level, these monitoring and control tasks are commonly performed using dedicated automation platforms specifically designed for continuous process operation. Pilot plants for the CO2 capture process (CO2-CP) typically manage operations using a programmable logic controller (PLC) or a distributed control system (DCS), both connected via Ethernet or Wi-Fi to a manufacturer-specific processor card linked to input/output modules, while large-scale capture plants require advanced control systems to promptly respond to load variations and minimize the time needed to return to steady-state operating conditions. [13], [14]. These cards connect directly to individual devices, such as temperature sensors, pressure transmitters, analyzers, flow transmitters, and control valves, for devices that operate with a 4–20 mA signal [15]. The choice between a PLC and DCS is influenced by the installation cost. It may be more practical to use one if only a small number of devices are needed, while the other may be more feasible or even preferable for larger installations [16]. Despite the complex of industrial automation systems, their high acquisition and maintenance costs often limit their implementation in laboratory-scale research facilities. In experimental process instrumentation, low-cost systems have emerged as effective and accessible alternatives for data acquisition and monitoring of key variables. This approach enables the replacement of expensive industrial equipment—such as CompactRIO modules or proprietary SCADA systems—with open-hardware platforms that provide sufficient accuracy and flexibility for academic and applied research purposes. [17]. Consequently, the adoption of low-cost instrumentation has become increasingly attractive for experimental research and engineering education. Laboratory-scale instrumentation, control, and automation are crucial to understanding thermal, mechanical, hydrological, and chemical processes, contributing to model predictions, as well as developing reliable tests, and reducing manual intervention [18], [19], [20]. In this context, the present work proposes a low-cost instrumentation and control platform for a laboratory-scale CO2 capture process, offering both practical and scientific contributions. The main contributions of this work are: Prediction of MEA saturation time during experimental tests under different solvent concentrations. Real time monitoring of the saturation curve during the CO2 capture process. The visibility of these types of processes at laboratory scale, and the importance of students having equipment instrumented and controlled in such a way that there is the least possible intervention, to obtain reliable data, since there is currently no literature on the instrumentation and control of such processes at laboratory scale. Implementation of a low-cost, real-time data acquisition system, based on calibrated sensors, enabling reliable measurements and providing high-quality datasets for subsequent data analysis and process optimization. To highlight the advantages of the proposed platform, Table 1 presents a comparison with another system implemented in the industrial sector:Table 1Comparative system.VariableIndustrial implementationCO2-CP SystemTemperatureResistance temperature detectors (RTDs), such as Pt100 and Pt1000, are known for their high accuracy but are expensive and require complex networking. [21]DS18B20 digital sensors (±0.5 °C) installed at strategic points in the systemThe data acquired from the CO2-CP System show it is working correctly, making it an excellent option for demonstrating the use of these technologies on a laboratory scale.FlowCertified industrial flowmeters with low latency. [14]Flowmeter Hall effect digital flowmeters, which operates measuring flow in the range of 50 to 800 mL/min.The installed Hall effect flowmeter allows liquid measurements in the laboratory, demonstrating the correct operation of the system. Table 1Comparative system.VariableIndustrial implementationCO2-CP SystemTemperatureResistance temperature detectors (RTDs), such as Pt100 and Pt1000, are known for their high accuracy but are expensive and require complex networking. [21]DS18B20 digital sensors (±0.5 °C) installed at strategic points in the systemThe data acquired from the CO2-CP System show it is working correctly, making it an excellent option for demonstrating the use of these technologies on a laboratory scale.FlowCertified industrial flowmeters with low latency. [14]Flowmeter Hall effect digital flowmeters, which operates measuring flow in the range of 50 to 800 mL/min.The installed Hall effect flowmeter allows liquid measurements in the laboratory, demonstrating the correct operation of the system. Comparative system. Hardware description The CO2-CP requires the instrumentation and control of several key variables for the study in question. Due to the number of sensors and control instruments, it was necessary to design and manufacture specific electronic boards to connect the sensors, to obtain a stable system during data acquisition and processing of the received signals. Fig. 1 shows the real complete system, highlighting the monitoring points of interest, as well as the location of the sensors and control instruments.Fig. 1Complete system with monitoring points. Complete system with monitoring points. For monitoring the CO2-CP, the following variables were instrumented: temperature, volumetric flow, and mass. Table 2 shows the nomenclature of the sensors instrumented in the process.Table 2Sensor nomenclature.Nomenclature SensorDescriptionTI-1Solution inlet temperature.TI-2Ambient temperature.TI-3Gas outlet temperature (0.70 m packed area).TI-4Solution outlet temperature (0.70 m packed area).TI-5Gas outlet temperature (0.25 m packed area).TI-6Solution outlet temperature (0.25 m packed area).TI-7Saturated solution temperature.TI-8Steam outlet temperature.TI-9Steam inlet temperature.TI-10Cooling fluid temperature.TI-11Regenerated solution outlet temperature.FI-3Digital hall effect flowmeter.EV-1Proportional solenoid valve.V13/2-way valve for airflow direction (inlet).V23/2-way valve for solution flow direction (inlet).V33/2-way valve for solution flow direction (outlet). Table 2Sensor nomenclature.Nomenclature SensorDescriptionTI-1Solution inlet temperature.TI-2Ambient temperature.TI-3Gas outlet temperature (0.70 m packed area).TI-4Solution outlet temperature (0.70 m packed area).TI-5Gas outlet temperature (0.25 m packed area).TI-6Solution outlet temperature (0.25 m packed area).TI-7Saturated solution temperature.TI-8Steam outlet temperature.TI-9Steam inlet temperature.TI-10Cooling fluid temperature.TI-11Regenerated solution outlet temperature.FI-3Digital hall effect flowmeter.EV-1Proportional solenoid valve.V13/2-way valve for airflow direction (inlet).V23/2-way valve for solution flow direction (inlet).V33/2-way valve for solution flow direction (outlet). Sensor nomenclature. Basic function of the temperature sensor For temperature monitoring at various points of interest, DS18B20-type digital temperature sensors were utilized. These sensors operate with a 5 V DC power supply and are waterproof, featuring a stainless-steel cover that encases their outer tip. Their operational range is −55 °C to + 125 °C, with an accuracy of ± 0.5 °C between −10 °C and + 85 °C and a configurable resolution of 9 to 12 bits. Additionally, they communicate using a single pin. The DS18B20 sensor was selected and implemented since the measurement intervals are within the appropriate temperatures for the currents in CO2-CP, such as the MEA, in the experimental tests the highest obtained temperature is approximately 27 °C. Furthermore, the highest value of saturated steam is approximately 95 °C. The electrical connection diagram for the digital temperature sensor is illustrated in Fig. 2.Fig. 2The electrical connection of the temperature sensor. The electrical connection of the temperature sensor. Basic function of the flowmeter For volumetric flow control of the absorbed solution, a commercial Hall effect digital flowmeter was used, which operates with an input voltage of 5 V DC and allows measuring flow in the range of 50 to 800 mL/min. This flowmeter works with small flows that occur in the system, which are favorable for the absorption column since the flows it handles are within the range of this component. In addition, a DN8-25 type V proportional ball solenoid valve was implemented, which operates with a voltage of 0 to 10 V DC. The opening and closing of the solenoid valve are controlled by a voltage regulator module, which adjusts the necessary voltage according to the degree of opening required to achieve the volumetric flow defined by the user. Fig. 3, Fig. 4 show the electrical diagrams of the digital flowmeter, the proportional solenoid valve, and the voltage regulator module used in the system.Fig. 3Electrical connection of the flowmeter.Fig. 4Electrical connection of the flow control system. Electrical connection of the flowmeter. Electrical connection of the flow control system. Basic function of the 3/2-way solenoid valve The post-combustion CO2 capture process employs three 3/2-way solenoid valves that operate at a voltage of 12 V DC together with a relay module that operates at 5 V DC. The valve system allows the flow of air current and absorbent solution to be directed between the two columns according to the configuration desired by the user. These valves are located at the inlets and outlets of the columns to direct the currents of Air, CO2, Monoethanolamine (MEA), and saturated MEA. Fig. 5 shows the electrical connection diagram of the solenoid valves together with the relay modules.Fig. 5Electrical connection of the 3/2-way solenoid valve. Electrical connection of the 3/2-way solenoid valve. Design files summary All designs are available in an editable format and archived in the Zenodo repository [1]. It shows all the documentation and designs implemented in the CO2-CP system. Table 3 shows a Design Files Summary.Table 3Design Files Summary CO2-CP system.FileDescriptionSectiontemperature.inoArduino code of one temperature sensor2.1temperature_ds18b20_datasheet.pdfTemperature sensor datasheet2.1flowmeter_proteus.pdsprjElectrical connection of the flowmeter2.2volumetric_flow_code.inoArduino code of the flowmeter2.2Voltage regulator module_datasheet.pdfVoltage regulator module datasheet2.2stream__select_code.inoArduino code to control one solenoid valve2.3stream_select.pdsprjElectrical connection of one solenoid valve2.3temperature_aborption.pdsprjElectrical connection of absorption temperature sensors5.3temperature_desorption.pdsprjElectrical connection of desorption temperature sensors5.3temperature_system_code.inoArduino code of both columns’ temperature sensors5.3stream_select_system_code.inoArduino code to control three solenoid valves5.6stream_select_system_simulation.pdsprjElectrical connection of three solenoid valve5.6PCB_1.cvr3dPCB 1 Aspire Files5.7PCB_2.cvr3dPCB 2 Aspire Files5.7PCB_1.gcodePCB 1 G code to manufacture5.7PCB_2.gcodePCB 2 G code to manufacture5.7PCB_1.pcbPCB 1 Printed Circuit Board files5.7PCB_2.pcbPCB 2 Printed Circuit Board files5.7CO2-CP system.pdfOperating manual system6 Table 3Design Files Summary CO2-CP system.FileDescriptionSectiontemperature.inoArduino code of one temperature sensor2.1temperature_ds18b20_datasheet.pdfTemperature sensor datasheet2.1flowmeter_proteus.pdsprjElectrical connection of the flowmeter2.2volumetric_flow_code.inoArduino code of the flowmeter2.2Voltage regulator module_datasheet.pdfVoltage regulator module datasheet2.2stream__select_code.inoArduino code to control one solenoid valve2.3stream_select.pdsprjElectrical connection of one solenoid valve2.3temperature_aborption.pdsprjElectrical connection of absorption temperature sensors5.3temperature_desorption.pdsprjElectrical connection of desorption temperature sensors5.3temperature_system_code.inoArduino code of both columns’ temperature sensors5.3stream_select_system_code.inoArduino code to control three solenoid valves5.6stream_select_system_simulation.pdsprjElectrical connection of three solenoid valve5.6PCB_1.cvr3dPCB 1 Aspire Files5.7PCB_2.cvr3dPCB 2 Aspire Files5.7PCB_1.gcodePCB 1 G code to manufacture5.7PCB_2.gcodePCB 2 G code to manufacture5.7PCB_1.pcbPCB 1 Printed Circuit Board files5.7PCB_2.pcbPCB 2 Printed Circuit Board files5.7CO2-CP system.pdfOperating manual system6 Design Files Summary CO2-CP system. Bill of materials summary Table 4 shows the necessary elements for the control and acquisition system of the post-combustion CO2 capture process variables at the laboratory scale. It also shows the electronic components and total cost of the CO2-CP system.Table 4Bill of materials and costs for CO2-CP system implementation.ComponentModelEmployed elementsCost per unit −currencyTotal cost −currencySource of materialsCodeOperating intervalMicroprocessorESP32 NodeMCU-32S2USD 9USD 18AmazonNA5 V DCTemperature sensorDS18B2011USD 3USD 33AmazonTI−55° C − 125° CPower supplyNANKADF1USD 80USD 80AmazonNAHall Effect Flowmetersensor TDS1USD 13.5USD 13.5AmazonFI50 – 800 mL/minVoltage regulator moduleDC Buck-Boost Converter1USD 46USD 46AmazonNA0 – 32 V DCSolenoid valveDN8-25 V-type ball solenoid valve1USD 75USD 75AliExpressEV0 – 10 V DC3/2-way solenoid valve2-position 3-way mini solenoid valve.3USD 30USD 90AmazonV12 V DCRelay Module2-channel relay module.2USD 6USD 12AmazonNA5 V DCResistancesElectrical resistance, 4.7kΩ11USD 18USD 18AmazonNANA Table 4Bill of materials and costs for CO2-CP system implementation.ComponentModelEmployed elementsCost per unit −currencyTotal cost −currencySource of materialsCodeOperating intervalMicroprocessorESP32 NodeMCU-32S2USD 9USD 18AmazonNA5 V DCTemperature sensorDS18B2011USD 3USD 33AmazonTI−55° C − 125° CPower supplyNANKADF1USD 80USD 80AmazonNAHall Effect Flowmetersensor TDS1USD 13.5USD 13.5AmazonFI50 – 800 mL/minVoltage regulator moduleDC Buck-Boost Converter1USD 46USD 46AmazonNA0 – 32 V DCSolenoid valveDN8-25 V-type ball solenoid valve1USD 75USD 75AliExpressEV0 – 10 V DC3/2-way solenoid valve2-position 3-way mini solenoid valve.3USD 30USD 90AmazonV12 V DCRelay Module2-channel relay module.2USD 6USD 12AmazonNA5 V DCResistancesElectrical resistance, 4.7kΩ11USD 18USD 18AmazonNANA Bill of materials and costs for CO2-CP system implementation. Build instructions Absorption and desorption process This section presents the electrical connection diagrams for each process variable, implementing the components listed in Table 4. The CO2-CP system consists of two main processes: absorption and desorption. Each process is represented by a block diagram, as shown in Fig. 6. The absorption process uses the following raw materials: monoethanolamine (MEA), the chemical absorbent; CO2, the solute to be absorbed from the gas stream; and air, used to dilute the solute to the required concentration for the experiment.Fig. 6Block diagram for CO2 −CP system description. Block diagram for CO2 −CP system description. The desorption process uses a monoethanolamine solution saturated with CO2 and saturated steam, which serve as inert gases to desorb the CO2 from the solution. The process variables monitored and controlled in both processes are temperature and volumetric flow rate. Finally, during absorption, MEA absorbs CO2 via an exothermic reaction, releasing energy and heating the absorbent solution. This process produces a saturated MEA solution containing the absorbed CO2. In the desorption process, the CO2 is separated from the saturated MEA to regenerate the absorbent and release the gas. This process releases CO2 as a gas, and the regenerated MEA is recirculated back into the absorption process. Description process CO2-CP The CO2-CP at a laboratory scale, situated in the Separation Processes Laboratory of the Unitary Operation Laboratory (LOU-FCQeI) at the Universidad Autónoma del Estado de Morelos (UAEM), is divided into two operational units: an absorption unit and a desorption unit. The CO2-CP comprises two packed columns; one is 0.076 m (3 in) in diameter and 0.70 m in length, made of 304 stainless steels, while the other column is 0.042 m (2 in) in diameter and 0.25 m in length, constructed from acrylic for the absorption process. Each column features two inlets and two outlets: one inlet at the top (liquid inlet) and one at the bottom (gas inlet), with one outlet located at the top (gas outlet) and another at the bottom (liquid outlet). Fig. 7 illustrates the complete system using a piping and instrumentation diagram (P&ID).Fig. 7Post-combustion process on a laboratory scale diagram. Post-combustion process on a laboratory scale diagram. The capture process starts at the bottom of the absorption columns, which are fed by a gas mixture of CO2-Air (streams 1 and 2) coming from an air compressor and a CO2 storage tank, passing through two flowmeters (Fl-1, Fl-2). The gas mixture enters a 3/2-way valve (V1) located at the bottom of the columns (stream 3). This valve directs the gas mixture towards a larger capacity absorption column (1600 mL) or a smaller capacity one (125 mL), streams 4 and 5 correspondingly. An absorbent solution located in a storage tank (MEA-H2O) is fed through the top of the absorption columns (stream 6) employing a peristaltic pump (E1). The supplied absorbent solution reaches a 3/2-way valve (V2), which allows the filling of the 1600 mL or 125 mL column (streams 7 and 8). The absorbent solution is put in countercurrent contact with the CO2-Air mixture through random packings, generating an exothermic reaction and increasing the temperature of the solution at the outlet. The CO2-saturated absorbent solution gets out of the bottom of the operating absorption column (stream 9) through a 3/2-way valve (V3). The saturated CO2 absorbent solution is released through stream 9, passing through a digital Hall effect flowmeter (FI-3) and a proportional solenoid valve (EV-1), arriving at a storage tank (TA-1). The CO2-free gas stream gets out to the top of the absorption column until it reaches a gas analyzer (stream 10). From the storage tank TA-1, the saturated solution obtained from the absorption process is pumped through the peristaltic pump (E1) to the desorption column (stream 11). Saturated steam (stream 13) from a steam generator enters the bottom of the desorption column, creating countercurrent contact between the fluids in the packed desorption column to regenerate the solution. After desorption, the regenerated solution is out (stream 12) at the bottom of the column and is stored in a storage tank (TA-2). Finally, water vapor and CO2 6 get out of the top of the column, passing through a heat exchanger. The condensate (stream 14) is stored in a storage tank (TA-3). For the development of the research work, the post-combustion CO2 capture system was fully implemented: a digital micro flowmeter by Hall effect (FI-3), a proportional solenoid valve (EV-1), a voltage regulator module, and 3 3/2-way solenoid valves at the inlet and outlet of the absorption and desorption columns respectively (V1, V2, V3). In addition, to monitor the temperature variable, the system was equipped with a total of 11 temperature sensors; two of them at the top and bottom of the 1600 mL absorption column (TI-3, TI-4), two at the top and bottom of the 125 mL absorption column (TI-5, TI-6), a sensor immersed in the absorbent solution storage tank (TI-1), one exposed to the environment (TI-2), one in the storage tank (TA-1) of the saturated absorbent solution (TI-7), two sensors at the top and bottom of the desorption column (TI-8, TI-9), a sensor located in the heat exchanger (TI-10) and another in the storage tank (TA-4) submerged in the regenerated solution (TI-11). Temperature Eleven temperature sensors were implemented at different points of interest in the post-combustion process for data acquisition. The sensors are divided into two phenolic plates, one for the absorption columns and one for the desorption column. Fig. 8 shows the electrical connections diagram of the 7 sensors located in the absorption columns, and Fig. 9 shows the electrical connections diagram of the 4 sensors located in the desorption column.Fig. 8Electrical connection temperature sensor for the desorption column.Fig. 9Electrical connection of temperature sensor for absorption columns. Electrical connection temperature sensor for the desorption column. Electrical connection of temperature sensor for absorption columns. Process calibration The calibration of the temperature sensors was performed using a thermal bath which contains a uniform fluid, and the temperature can be adjusted. The test was made at different temperatures which were compared with the temperature acquired by the CO2-CP system sensors, established by reference the thermal bath. Once the sensors have been installed and connected, the data acquisition system is calibrated and adjusted. Calibration involves comparing measurements obtained from the instrument under evaluation with those from a reference standard to quantify and correct any deviations in the recorded values. To perform this procedure, a thermal bath is employed, providing a controlled, uniform temperature environment that can be maintained at stable operating conditions across a range of setpoints. The sensors are first identified, then immersed in the thermal bath fluid to record measurements under each test condition. The calibration process begins at 5 °C and continues in successive increments of 5 °C until reaching 115 °C. At each temperature setpoint, the sensor readings are recorded and compared with the reference values. This procedure enables the determination of the calibration and correction equations needed to improve measurement accuracy and reliability during experimental operation. The coefficient of determination (R2) indicates how well the calibration equation represents the experimental data. Its value ranges from 0 to 1, with values closer to 1 indicating a better fit. R2 is calculated using Equation (1):(1)R2=1-∑yi-y^i2∑yi-y¯2where (yi) is the reference temperature, (y^i) is the temperature estimated by the calibration equation, and (y¯) is the mean value of the reference temperatures. The Root Mean Square Error (RMSE) quantifies the standard deviation of the residual errors after applying the calibration correction. An RMSE value of approximately 0.05 °C is generally considered excellent; values around 0.20 °C are considered good, whereas values greater than 0.5 °C are relatively high for calibrated temperature sensors. RMSE is calculated using Equation (2):(2)RMSE=1n∑i=1nyi-y^i2where the error term is defined by Equation (3):(3)ei=yi-y^iThe Mean Absolute Error (MAE) represents the average magnitude of the errors regardless of their sign. It provides a direct measure of the average deviation between the measured and reference values and is calculated using Equation (4):(4)MAE=1n∑i=1nyi-y^iThe Maximum Error corresponds to the largest absolute deviation recorded throughout the entire calibration procedure and is determined using Equation (5):(5)Errormax=maxeiFinally, the Corrected Accuracy, which reflects the measurement accuracy after applying the calibration correction, is calculated using Equation (6):(6)CA=100-MAE∗100%Table 5 shows the correction equations for the most interesting temperature sensor in the CO2-CP system.Table 5Correction equations of temperature sensor.SensorCorrection EquationsR2(−)RMSE(−)MAE(°C)Max AE (°C)CA(%)TI-3TI3=0.9932TI1+1.05090.997640.005090.829311.02495.25TI-4TI4=0.9897TI2+0.58700.999670.001800.252470.66097.82TI-5TI5=0.8420TI5+1.37770.997230.008640.865751.37794.34TI-6TI6=0.9932TI6+0.35470.999890.000520.144840.35598.79TI-7TI7=0.9848TI7+1.02260.997610.005180.834911.03595.21TI-8TI8=0.9693TI8+1.60350.999660.001780.257230.65697.81TI-9TI9=0.9911TI9+0.76780.997220.008660.868501.37994.33TI-11TI11=0.9938TI11+0.35250.999890.000550.145610.36498.77 Table 5Correction equations of temperature sensor.SensorCorrection EquationsR2(−)RMSE(−)MAE(°C)Max AE (°C)CA(%)TI-3TI3=0.9932TI1+1.05090.997640.005090.829311.02495.25TI-4TI4=0.9897TI2+0.58700.999670.001800.252470.66097.82TI-5TI5=0.8420TI5+1.37770.997230.008640.865751.37794.34TI-6TI6=0.9932TI6+0.35470.999890.000520.144840.35598.79TI-7TI7=0.9848TI7+1.02260.997610.005180.834911.03595.21TI-8TI8=0.9693TI8+1.60350.999660.001780.257230.65697.81TI-9TI9=0.9911TI9+0.76780.997220.008660.868501.37994.33TI-11TI11=0.9938TI11+0.35250.999890.000550.145610.36498.77 Correction equations of temperature sensor. In the proposed monitoring and instrumentation system, the DS18B20 temperature sensor was selected due to its low cost, wide operating range, and ease of integration with the ESP32 microcontroller. This digital device uses a 1-Wire communication interface, allowing multiple sensors to be connected to a single data pin, thereby reducing wiring complexity and eliminating the need for signal conditioning circuits. Unlike PT100 RTD sensors, which require analog compensation circuits and periodic calibration to ensure accuracy, the DS18B20 provides direct digital readings, simplifying data acquisition and improving the system’s modularity. Furthermore, its significantly lower cost makes it an ideal option for research, educational, or automation platforms focused on low-cost instrumentation, where affordability and simplicity are key design considerations. Table 6 shows a comparative analysis between both sensors, highlighting the advantages of the DS18B20 in terms of cost, connection simplicity, and applicability in experimental processes instrumented through the ESP32 platform.Table 6Comparative temperature sensors.CriterionDS18B20 SensorPT100 (RTD) SensorComments/PreferenceSignal typeDigital (1-Wire)Analog (resistance variation)The DS18B20 eliminates the need for analog conditioning, simplifying the design.Measurement range–55 °C to + 125 °C–200 °C to + 850 °CThe PT100 has a wider range, but the DS18B20 adequately covers most laboratory and experimental processes.Resolution / Accuracy12-bit (±0.5 °C typical)High precision (±0.1 °C or better)The PT100 provides higher accuracy, although the DS18B20 is sufficient for experimental and educational measurements.Calibration requiredNo complex calibration requiredPeriodic calibration requiredAdvantage for the DS18B20 in educational and quick test systems.Ease of connection1-Wire digital interface (single data pin, multiple sensors on shared bus)Requires Wheatstone bridge or signal conditioning circuitThe DS18B20 offers direct integration with microcontrollers (Arduino, ESP32).Approximate cost3–5 USD25–40 USD (without conditioning)Major advantage of the DS18B20 for low-cost instrumentation projects.Power supply3.0–5.5 VDCStable excitation current (100 μA–1 mA)The DS18B20 is simpler in design and power requirements.Communication / readingDirect digital, 1-Wire protocolAnalog conversion using external ADC or conditioning moduleThe DS18B20 reduces the need for additional hardware.Typical applicationsThermal monitoring in experiments, automation, IoTHigh-precision industrial processes, quality controlThe DS18B20 is ideal for experimental and educational instrumentation. Table 6Comparative temperature sensors.CriterionDS18B20 SensorPT100 (RTD) SensorComments/PreferenceSignal typeDigital (1-Wire)Analog (resistance variation)The DS18B20 eliminates the need for analog conditioning, simplifying the design.Measurement range–55 °C to + 125 °C–200 °C to + 850 °CThe PT100 has a wider range, but the DS18B20 adequately covers most laboratory and experimental processes.Resolution / Accuracy12-bit (±0.5 °C typical)High precision (±0.1 °C or better)The PT100 provides higher accuracy, although the DS18B20 is sufficient for experimental and educational measurements.Calibration requiredNo complex calibration requiredPeriodic calibration requiredAdvantage for the DS18B20 in educational and quick test systems.Ease of connection1-Wire digital interface (single data pin, multiple sensors on shared bus)Requires Wheatstone bridge or signal conditioning circuitThe DS18B20 offers direct integration with microcontrollers (Arduino, ESP32).Approximate cost3–5 USD25–40 USD (without conditioning)Major advantage of the DS18B20 for low-cost instrumentation projects.Power supply3.0–5.5 VDCStable excitation current (100 μA–1 mA)The DS18B20 is simpler in design and power requirements.Communication / readingDirect digital, 1-Wire protocolAnalog conversion using external ADC or conditioning moduleThe DS18B20 reduces the need for additional hardware.Typical applicationsThermal monitoring in experiments, automation, IoTHigh-precision industrial processes, quality controlThe DS18B20 is ideal for experimental and educational instrumentation. Comparative temperature sensors. Volumetric flow The volumetric flow in the absorption columns was instrumented and controlled to operate the system continuously. A solenoid valve and a voltage regulator module were employed to regulate the volumetric flow, while a digital Hall-effect flowmeter was utilized for flow measurement. The volumetric flow rate for operating the absorption columns depends on the voltage supplied to the solenoid valve. Subsection 2.2 shows the electrical connections of the elements to the instrument and control the volumetric flow variable. Stream select The post-combustion CO2 capture process involves two absorption columns with different capacities. Therefore, three 3/2-way solenoid valves were installed, allowing the configuration of which column to operate during experimental testing. The valves are located at the columns' inlet and outlet streams. Fig. 10 shows the electrical connection diagram of the 3 3/2-way solenoid valves implemented in the system.Fig. 10Electrical connection of the 3/2-way solenoid valves. Electrical connection of the 3/2-way solenoid valves. Internal wiring Due to the specific work conditions and quantity of instruments, two printed circuit boards (PCBs) were manufactured for the electrical connections for every instrument. The PCB1 board was designed for the absorption columns of the post-combustion system. This PCB, shown in Fig. 11, contains 4 temperature sensors, 1 digital flowmeter, 1 proportional solenoid valve, 1 voltage regulator module, and 2 double relay modules, considering their voltage supply at 12 and 5 V, in addition to the respective signals of each of the elements.Fig. 11Manufactured design of PCB1. (Nomenclature in Table 2). Manufactured design of PCB1. (Nomenclature in Table 2). The PCB2 board was designed for the desorption column of the post-combustion system. This PCB, shown in Fig. 12, contains 7 temperature sensors, as well as additional connections for future instrumentation.Fig. 12Manufactured design of PCB2. (Nomenclature in Table 2). Manufactured design of PCB2. (Nomenclature in Table 2). The PCBs were manufactured and assembled, integrating the modules and connections necessary for the instrumentation and control of the process. Fig. 13 shows the electrical control panel for the process. The panel has two ESP32 cards (1–2), one assigned to the absorption column sensors (3), and the other to the desorption column sensors (4).Fig. 13CO2-CP system control cabinet. CO2-CP system control cabinet. Operation instructions Consult the repository where the CO2-CP System operating manual was added [1]. Validation and characterization In this section, the correct operation of the instrumentation and control of the process variables was validated. Table 7 shows the operating conditions for the post-combustion CO2 capture process at the laboratory scale.Table 7General operating parameters.VariableParametersSystem pressure0.8 atmSteam generator pressure4.5 atmAmbient temperature20 °CRelative humidity50 – 65%Gas flow rate (air)2 – 10 L/minGas flow rate (CO2)0.1 – 1 L/minLiquid stream feed flow10 – 40 ml/minMEA concentration10 – 30 %w/wCO2 concentration at the gas stream inlet5 – 12.5 %v/v Table 7General operating parameters.VariableParametersSystem pressure0.8 atmSteam generator pressure4.5 atmAmbient temperature20 °CRelative humidity50 – 65%Gas flow rate (air)2 – 10 L/minGas flow rate (CO2)0.1 – 1 L/minLiquid stream feed flow10 – 40 ml/minMEA concentration10 – 30 %w/wCO2 concentration at the gas stream inlet5 – 12.5 %v/v General operating parameters. Temperature variable The correct operation of sensors TI-1 − TI-6 was validated through experimental saturation tests, which consist of measuring the temperature of the process streams at the inlets and outlets of the absorption columns. Fig. 14 shows the temperatures acquired during the absorption tests performed in the smaller column, the TI-1 and TI-2 are kept at ambient temperature since they are located at the inlets of the column, sensors TI-5 and TI-6 measured the heat released during absorption, thereby confirming the exothermic reaction. This behavior was reflected by an increase in the temperature of the absorbent solution as it reacted with the CO2 flowing in countercurrent. To verify that this thermal behavior was independent of the column size, the same experimental procedure was subsequently applied to the larger absorption column. Similarly, experimental tests were carried out on the large absorption column, sensors TI-3 and TI-4 exhibited temperature profiles comparable to those recorded by the outlet sensors of the smaller column, demonstrating the repeatability and consistency of the absorption process under different operating configurations.Fig. 14Temperature sensors in the smaller absorption column (0.25 m packed area). Temperature sensors in the smaller absorption column (0.25 m packed area). After validating the instrumentation of the absorption stage, the performance of the sensors installed in the regeneration stage was evaluated. The validation TI-7 − TI-11 sensors was carried out using an experimental desorption test to regenerate the absorbent solution used in the CO2 capture process. Saturated steam generated by the steam generator was employed as the heat source. The temperature profiles obtained during the desorption process confirmed the proper operation of the monitoring system. Fig. 15 shows the constant temperature of the saturated absorbent solution environment. Sensor T11 registers a slight increase in the cooling fluid temperature because the fluid is recirculating in the same storage tank. Likewise, sensors T8, T9, and T10 show an increase in temperature until reaching the steady state, corresponding to the saturation temperature of the steam used in the process.Fig. 15Temperature sensors in the desorption column. Temperature sensors in the desorption column. Volumetric flow variable The volumetric flow of the absorbent solution (MEA-H2O) was instrumented and controlled. For this purpose, an EV-1 solenoid valve was used; the opening and closing process was controlled by a voltage regulator module that supplied the necessary voltage. In addition, a FI-3 digital Hall effect flowmeter and a TI-6 temperature sensor were integrated at the outlet of the saturated solution. The countercurrent contact between the absorbent solution and the CO2 stream inside the absorption column promotes an exothermic reaction, demonstrating an increase in the temperature of solution [18]. Fig. 16 shows the temperature profile of the absorbent solution as a function of time for a 10 % MEA concentration.Fig. 16Validation of the continuous process (EV-1, FI-3). Validation of the continuous process (EV-1, FI-3). The implementation of the temperature sensors implemented in the CO2-CP system allows the temperature variable to be monitored in real time until it reaches a steady state in the various tests performed. In addition to temperature monitoring, maintaining an appropriate flow balance is essential for ensuring stable process operation. On the other hand, the implementation of the flow control system ensures compliance with the continuity principle, which states that the volumetric flow rate entering the absorption column equals the volumetric flow rate leaving the system under steady-state conditions. From a hardware integration perspective, the developed electronic architecture also contributes to improving the overall reliability of the data acquisition system. The incorporation of dedicated printed circuit boards (PCB) improves the organization of measurement and control devices, enhances signal stability, and reduces electrical noise, thereby enhancing the reliability and quality of the acquired experimental data. Conclusions With respect to the previous reports LOU configuration, Fig. 17, Fig. 18 compare the system’s behavior before and after the implementation of the proposed instrumentation and control platform. This comparison clearly illustrates the improvements achieved in data acquisition and process monitoring during experimental tests conducted under different operating conditions.Fig. 19Fig. 20.Fig. 17Data acquisition before instrumentation and control.Fig. 18Data acquisition before instrumentation and control.Fig. 19System before implementation of the instrumentation and control.Fig. 20System after implementation of the instrumentation and control. Data acquisition before instrumentation and control. Data acquisition before instrumentation and control. System before implementation of the instrumentation and control. System after implementation of the instrumentation and control. Real-time temperature monitoring enables the user to verify that the CO2 captures process is operating correctly, while simultaneously acquiring reliable experimental data for subsequent analysis. This capability not only facilitates process supervision but also enables early identification of abnormal operating conditions, thereby improving the safety and reliability of the experimental system. In addition to temperature monitoring, stable flow regulation is essential to ensure continuous operation of the absorption process. Consequently, the implementation of volumetric flow control prevents the absorption column from either flooding or running dry, significantly reducing operator intervention during experimental testing. To further improve the usability of the proposed platform, a dedicated graphical interface was developed for real time visualization and process management. The implementation of this low-cost data acquisition system enables users to continuously monitor operating variables during experiments. In addition to meeting system requirements, such as installing sensors at strategic locations, the flow control system enables the operator to select the absorption column based on the experimental conditions being investigated. The overall impact of the proposed instrumentation strategy is illustrated in figures 19 and 20, which present the CO2-CP system before and after the implementation of the instrumentation and control platform. To ensure the reliability of the acquired measurements, all temperature sensors were calibrated prior to the experimental evaluation. The calibration results show agreement between the corrected and reference temperatures for all evaluated sensors. The coefficient of determination (R2) ranged from 0.99722 to 0.99989 units, indicating a linear correlation and confirming the suitability of the proposed calibration equations. The Mean Squared Relative Error (MSRE) values remained below 0.0009 °C, while the Mean Absolute Error (MAE) ranged from 0.145 °C to 0.869 °C. Likewise, the Maximum Absolute Error (Max AE) was lower than 1.38 °C for all sensors. Overall, the statistical indicator confirms that the proposed calibration methodology significantly improved the accuracy and reliability of the temperature measurements, providing high quality experimental data for subsequent thermal analysis and process evaluation. The corrected accuracy (CA) remained above 94% for all calibrated sensors. Ethics statements No human or animal studies were conducted in the design of this work. CRediT authorship contribution statement María Fernanda Silva León: Writing – original draft, Methodology, Conceptualization. Erick Omar Castañeda Magadán: Writing – review & editing, Writing – original draft, Supervision, Investigation, Data curation. Miriam Navarrete Procopio: Writing – review & editing, Writing – original draft, Validation. Víctor Manuel Zezatti Flores: Writing – review & editing, Writing – original draft, Visualization, Supervision, Investigation. Ángel Tlatelpa Becerro: Supervision. Declaration of competing interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.